Trumpet

REAL EAR MEASUREMENT SYSTEM

Trumpet is an advanced and easy-to-use Real Ear Measurement system, capable of performing quick and accurate in-situ measures thanks to its small lightweight probes and its built-in amplified speaker. Trumpet also incorporates a diagnostic audiometer for carrying out air, bone and speech audiometric exams.
Controlled directly from your computer, Trumpet offers a complete but simple and intuitive software interface, making it a compact and portable system with no compromises.

FITTING SYSTEM

As a fitting system, Trumpet is capable of measuring all the standard gain/response values: unaided (REUG/R), aided (REAG/R), insertion (REIG) and occluded (REOG/R).
Sequences of measurements can be set and run with just a few clicks, making the daily workflow that much quicker and easier.

TARGET PROTOCOLS

Trumpet offers the most recent target protocols, like NAL-NL1, NAL-NL2 and DSL v5, to assist acousticians in the setting of hearing aid parameters, so that fittings will be well accepted, and patients less likely to return dissatisfied.

LIVE MODE FOR SPEECH MAPPING

This most recent and effective counseling tool is naturally available on Trumpet. Live mode for speech mapping allows you to demonstrate the benefit of wearing the hearing aid straight away, by simply talking to the patient, just as any family member might do.

ADVANCED MODE

Input-output view mode, higher frequency resolution and complete user customizable work procedure are just some of the additional features available in Advanced mode.

COUPLER KIT for RECD

With Trumpet, there is also the option of a kit with built-in speaker, 2cc coupler and insert earphones, which will allow RECD measures to be taken from non-collaborative patients without the need for a hearing instrument test-box.

DIAGNOSTIC AUDIOMETER

In audiometer mode, Trumpet can perform air, bone, free field and speech audiometry exams, Hughson-Westlake automated audiometry, QuickSIN®, and includes a 2 channel Master Hearing Aid.

COMPUTER CONTROL

Connect Trumpet to the USB port of your Windows® computer and you are ready to go: no driver and no additional cable required.
You can control Trumpet through the amazingly user-friendly Maestro software interface and, if need be, transfer the data directly to Noah.

Download Technical Sheet

REM system

REM standardEC 61669, ANSI S3.46.
Stimuli (REM)White noise, Pink noise, Speech noise, ICRA, Real Speech, any other sound file
Frequency range 125 Hz – 12 kHz
Intensity range50 – 90 dB SPL
Frequency Resolution1/3, 1/24 octave
Available testsREUR / REUG, REAR / REAG, REIG, REOR / REOG, Advanced, Live speech, RECD (opt.)

Audiometer

IntensityAC: -10 to 120 dB HL, BC: -10 to 80 dB HL
Frequency rangeAC: 125 – 8000 Hz, BC: 250 - 8000 Hz
Stimulipure tone, warble tone
Speech inputsUSB, Mic
FF outputYes (built-in amp.)
Test battery PTA (HL and UCL), Speech Audiometry (WRS, SRT, MCL, and UCL), Autothreshold, Master Hearing Aid, QuickSIN® (opt.).
StandardsAudiometer: EN 60645 -1 / ANSI S3.6, Type 2, Speech: EN60645-2 / ANSI S3.6 type A-E
Connection to PCUSB (driverless)
Compatible SWMaestro software
Noah compatibilityYes
Dimensions15.5 x 10 x 24.5 cm / 6.1 x 3.9 x 9,7 in (WxDxH)
Weight 1.5 kg / 3.5 lb
SafetyEN 60601-1
EMCEN 60601-1-2